Artificial joints operate in the mixed lubrication regime, not with the full fluid films observed for natural human joints under physiological loading. As a result, artificial joints will experience higher friction and wear due to metal-plastic contact. Full fluid film lubrication is achieved in the natural joint as a result of the high degree of compliance of the joint surfaces. One way of improving the performance of the artificial joint is to introduce a more compliant surface.
Measuring the joint friction and plotting the resulting Stribeck curve (the relationship between friction coefficient and the parameter velocity x viscosity x radius/load) is of fundamental interest in the design and development of better artificial joints.
Test machines have been developed at Durham University by Professor Tony Unsworth to study the frictional performance of hip and knee joint prostheses (see references at the end of this document). These machines operate with a simplified on/off loading cycle compared with physiological loading, in order to provide steady-state friction measurements during the loading phase for Stribeck analysis.
Plint & Partners Ltd have manufactured a modified version of the Durham Friction Simulators, with the agreement of Professor Unsworth. Comparative tests are being performed to ensure correlation with data generated on the original Durham machines.
The key performance requirements for the Friction Simulator are that it applies a defined load cycle in a predictable and repeatable fashion and measures the very low levels of friction generated within the lubricated joint.
The machine comprises two support systems for the hip and knee joint components. Hip joints are inverted from their natural position. The acetabular cup or tibial plate component is mounted in a low friction trunnion mounted reservoir carried in a loading frame. The femoral component (either a taper mounted ball or special fixed axis knee piece) is mounted on the underside of a rigid cross-beam that is oscillated through an adjustable pre-set angle. A jig is provided to ensure that the femoral component is oscillated about its centre and is aligned with the lower test piece.
A high duty linear solenoid acting through a 9:1 lever applies the load to the carriage that supports the trunnion mounted reservoir. This system is compact, stiff and provides an easily controllable loading cycle. The load is measured by a piezo-electric force transducer mounted centrally on the carriage.
The trunnion bearings are mounted in such a way as to permit their inner races to be driven. Rotating the bearings increases their load carrying capacity from their static to their dynamic rating and reduces the parasitic friction from static to rolling friction. By driving the bearings in opposite directions the parasitic torque transmitted to the trunnion assembly by the bearings substantially cancels out. A parasitic torque of about 10% of the smallest measured friction torque is expected.
The trunnion mounted assembly is restrained from rotation by a piezo-electric force transducer carried on the lower loading carriage and connected by means of a flexure link.
Two reservoirs are provided, one for the acetabluar cup and one for the tibial plate. Tests may be run dry and lubricated with water, synthetic lubricants and synovial fluid.
The cross beam carrying the femoral component is carried in bearings that are in line with the axis of the lower trunnion mounted assembly. These bearings are carried on brackets mounted on linear ball slides which allow the lower specimen assembly to be located correctly and locked in position.
The beam is oscillated sinusoidally through a crank mechanism with an adjustable throw. The mass of the arm and crank mechanism are minimised in order to minimise out of balance forces and thus to reduce the potential for parasitic excitation of other parts of the assembly. The drive arrangement comprises a thyristor controlled dc gear motor with tachogenerator feedback.
The timing between the load on/off action and the oscillating action is adjusted by means of an optical sensor and adjustable trigger. This permits the position of load and the duration of load application to be adjusted. It also permits the load to be applied on either the forward or reverse strokes.
The has PC based sequence programmable control and data acquisition. This is provided by an integrated Serial Link Interface Module and COMPEND 2000 software running on a host PC, operating under Windows. Data is stored to hard disc in standard spread sheet compatible file formats (.csv or .tsv).
Tests are defined by a sequence of steps, each step containing set-point, data recording rates and alarm level information. Set-points may be adjusted by step change or ramp. The test sequence is followed unless interrupted by the operator or an alarm. Set-points may also be adjusted manually using on screen toggles.
Load Range: 200 to 2,500 N
Load Sensor: piezo-electric transducer
highest sensitivity: 10 mN
highest range: 7,500 N
Frequency Range: 0.02 to 1 Hz
Oscillating Angle: 0 to +/- 35 degrees
Friction Sensor: piezo-electric transducer
highest sensitivity: 1 mN
highest range: 500 N
Maximum Knee Size: 85 mm wide
Interface: PLINT SLIM 2000 Serial Link Interface Module
Software: PLINT COMPEND 2000 Windows based sequence control and data acquisition software
High Speed DAQ: PLINT SLIM 2000 Serial Link Interface Module
Total Memory: 640 kBytes
Sample Size: 2 Bytes per channel per sample
Number of Channels: 1 to 8
Data Rates: 75 Hz, 150 Hz, 300 Hz, 600 Hz, 1200 Hz
Applied Load
Rotational Speed (Frequency)
Number of Cycles
Test Duration
Applied Load
Rotational Speed (Frequency)
Frictional Torque
Number of Cycles
Test Duration
Example publications on the original Durham Friction Simulators:
Unsworth A., Pearcy M. J., White E. F. T. and White G., Engineering in Medicine, 17 (30), 1988, 101-104.
Hall R. M., Unsworth A., Wroblewski B. M. and Burgess I. C., Wear, 175, 1994, 159-166.
| Electricity: | 220/240V, single phase, 50 Hz, 2.8 kW |
| 110/120 V, single phase, 60 Hz, 2.8 kW | |
| PC and Printer: | Minimum Specification |
Bench-mounting machine: 580 mm wide x 560 mm deep x 560 mm high
Bench-mounting cabinet: 540 mm wide x 540 mm deep x 450 mm high
Packing Specifications: 1.73 m3, GW 270 kg, NW 230 kg